지식나눔

refractive index detector (RID) 원리

HPLC 를 이용한 베타-agarase , D-galactose 분석을 하고자 합니다. 이때 LC 에 사용하는 RID 의 시료검출 원리와 참고자료를 구하고 있습니다. 구할수 있는 곳이라도 알려주시면 감사하겠습니다. 수고하십시요..
  • RID
  • refractive index detector
  • HPLC
지식의 출발은 질문, 모든 지식의 완성은 답변! 
각 분야 한인연구자와 현업 전문가분들의 답변을 기다립니다.
답변 2
  • 답변

    신정규님의 답변

    http://hplc.chem.shu.edu/BOOK/html/detector/det_ri.html 에 가시면 있습니다. Principles The refractive index (RI) detector is the only universal detector in HPLC. The detection principle involves measuring of the change in refractive index of the column effluent passing through the flow-cell. The greater the RI difference between sample and mobile phase, the larger the imbalance will become. Thus, the sensitivity will be higher for the higher difference in RI between sample and mobile phase. On the other hand, in complex mixtures, sample components may cover a wide range of refractive index values and some may closely match that of the mobile phase, becoming invisible to the detector. RI detector is a pure differential instrument, and any changes in the eluent composition require the rebalansing of the detector. This factor is severely limiting RI detector application in the analyses requiring the gradient elution, where mobile phase composition is changed during the analysis to effect the separation. Two basic types of RI detectors are on the market today. Both require the use of a two-path cell where the sample-containing side is constantly compared with the non-sample-containing reference side. 그 외 참고 사이트 http://www.bischoff-chrom.de/?url=katalog/browse&show=detektor www.postnova.com/PDFDocuments/ productinfos/RI_PN_3120_E_291003.pdf > >HPLC 를 이용한 베타-agarase , D-galactose 분석을 하고자 합니다. > >이때 LC 에 사용하는 RID 의 시료검출 원리와 참고자료를 구하고 있습니다. > >구할수 있는 곳이라도 알려주시면 감사하겠습니다. > >수고하십시요.. > > >
    http://hplc.chem.shu.edu/BOOK/html/detector/det_ri.html 에 가시면 있습니다. Principles The refractive index (RI) detector is the only universal detector in HPLC. The detection principle involves measuring of the change in refractive index of the column effluent passing through the flow-cell. The greater the RI difference between sample and mobile phase, the larger the imbalance will become. Thus, the sensitivity will be higher for the higher difference in RI between sample and mobile phase. On the other hand, in complex mixtures, sample components may cover a wide range of refractive index values and some may closely match that of the mobile phase, becoming invisible to the detector. RI detector is a pure differential instrument, and any changes in the eluent composition require the rebalansing of the detector. This factor is severely limiting RI detector application in the analyses requiring the gradient elution, where mobile phase composition is changed during the analysis to effect the separation. Two basic types of RI detectors are on the market today. Both require the use of a two-path cell where the sample-containing side is constantly compared with the non-sample-containing reference side. 그 외 참고 사이트 http://www.bischoff-chrom.de/?url=katalog/browse&show=detektor www.postnova.com/PDFDocuments/ productinfos/RI_PN_3120_E_291003.pdf > >HPLC 를 이용한 베타-agarase , D-galactose 분석을 하고자 합니다. > >이때 LC 에 사용하는 RID 의 시료검출 원리와 참고자료를 구하고 있습니다. > >구할수 있는 곳이라도 알려주시면 감사하겠습니다. > >수고하십시요.. > > >
    등록된 댓글이 없습니다.
  • 답변

    전주홍님의 답변

    첨부한 자료가 좋은 자료가 될지 모르겠네요... HPLC system관련 서적이 많이 나와있는데 간략하게는 설명이 되어있습니다. 혹시 사용하시는 기기의 회사 manual을 참고해 보셨는지요... > >HPLC 를 이용한 베타-agarase , D-galactose 분석을 하고자 합니다. > >이때 LC 에 사용하는 RID 의 시료검출 원리와 참고자료를 구하고 있습니다. > >구할수 있는 곳이라도 알려주시면 감사하겠습니다. > >수고하십시요.. > > >
    첨부한 자료가 좋은 자료가 될지 모르겠네요... HPLC system관련 서적이 많이 나와있는데 간략하게는 설명이 되어있습니다. 혹시 사용하시는 기기의 회사 manual을 참고해 보셨는지요... > >HPLC 를 이용한 베타-agarase , D-galactose 분석을 하고자 합니다. > >이때 LC 에 사용하는 RID 의 시료검출 원리와 참고자료를 구하고 있습니다. > >구할수 있는 곳이라도 알려주시면 감사하겠습니다. > >수고하십시요.. > > >
    등록된 댓글이 없습니다.